Command reference
R&S
®
ZNB/ZNBT
1246User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire......................................................1261
[SENSe<Ch>:]CORRection:POWer<PhyPt>:AWAVe[:STATe]........................................... 1262
[SENSe<Ch>:]CORRection:POWer:DATA...................................................................... 1262
[SENSe<Ch>:]CORRection:POWer:DATA:PORT<PhyPt>................................................ 1262
[SENSe<Ch>:]CORRection:POWer<PhyPt>:IMODulation:ACQuire...................................1263
[SENSe<Ch>:]CORRection:POWer<PhyPt>[:STATe]....................................................... 1264
[SENSe<Ch>:]CORRection:SMATrix:CDATa................................................................... 1265
[SENSe<Ch>:]CORRection:SMATrix:CDATa:PORT<PhyPt>.............................................1265
[SENSe<Ch>:]CORRection[:STATe]...............................................................................1266
[SENSe<Ch>:]CORRection:STIMulus?.......................................................................... 1266
[SENSe<Ch>:]CORRection:STIMulus:PORT<PhyPt>?.................................................... 1266
[SENSe<Chn>:]CORRection:PSTate?............................................................................1266
[SENSe<Chn>:]CORRection:SSTate?............................................................................1267
[SENSe<Ch>:]CORRection:CDATa <ErrorTerm>, <SourcePort>, <LoadPort>,
<CorrectionData>...
[SENSe<Ch>:]CORRection:CDATa:PORT<PhyPt> <ErrorTerm>, <SourcePort>,
<LoadPort>, <CorrectionData>...
Writes or reads system error correction data for a specific channel <Ch>, calibration
method ([SENSe<Ch>:]CORRection:COLLect:METHod:DEFine), and port combi-
nation <SourcePort>, <LoadPort>. The setting command can be used to transfer
user-defined correction data <CorrectionData> to the analyzer; the query returns
the current correction data set. ASCII or block data can be transferred, depending on
the selected data transfer format (FORMat[:DATA]).
The sweep must be stopped to transfer calibration data; see program example for
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected:DEFault.
Note: This command affects the active calibration of channel no. <Ch> or the factory
calibration (if no channel calibration is active). For the factory calibration, the query
form is allowed only (no change of factory calibration data).
Tip: The analyzer provides a default calibration corresponding to a test setup which
does not introduce any systematic errors; see [SENSe<Ch>:]CORRection:
COLLect:SAVE:SELected:DEFault.
For an overview of calibration methods and error terms refer to Chapter 4.5.1, "Calibra-
tion types", on page 163.
G and H matrices
The 7-term calibration types named Txx (e.g. TOM, TRM, TRL, TNA) are based on a
network analyzer with two ports i and j, each equipped with a test receiver and a refer-
ence receiver. The system errors are described in terms of two "error two-ports" P
G
and P
H
:
●
The error two-port P
G
is assigned to port i of the analyzer. Its transmission matrix G
describes how the system errors modify the outgoing and incident waves at port i:
test i
ref i
2221
1211
i
i
m
m
GG
GG
a
b
SCPI command reference