Command reference
R&S
®
ZNB/ZNBT
854User Manual 1173.9163.02 ─ 62
Query parameters:
<Format> SDATa
Read unformatted sweep data (fixed parameter): Returns the
real and imaginary part of each measurement point (2 values
per trace point, irrespective of the selected trace format).
<RvCount> Number of sweep to be read. 1 denotes the last sweep acquired,
2 denotes the second-last and so forth.
Range: 1 to sweep count defined via
[SENSe<Ch>:]SWEep:COUNt
Example:
SWE:COUN 10
Define the number of sweeps (10) to be measured in single
sweep mode.
INIT:CONT OFF; :INIT; *OPC?
Activate single sweep mode and start a single sweep sequence
in channel no. 1. Wait until the single sweep sequence is com-
plete.
CALC:DATA:NSW? SDAT,3
Query the results of the 8
th
sweep.
See also Chapter 8.2.4.3, "Retrieving the results of previous
sweeps", on page 1513.
Usage:
Query only
CALCulate<Ch>:DATA:SGRoup? <Format>
Reads the current response values of all S-parameters associated to a group of logical
ports (S-parameter group). The S-parameter group must be created before using
CALCulate<Ch>:PARameter:DEFine:SGRoup.
Suffix:
<Ch>
.
Channel number of the previously defined S-parameter group.
Query parameters:
<Format> FDATa | SDATa | MDATa
Output format for the S-parameter data, see
CALCulate<Chn>:DATA on page 846.
Example: See CALCulate<Ch>:PARameter:DEFine:SGRoup
Usage: Query only
CALCulate<Chn>:DATA:STIMulus?
Reads the stimulus values of the active data or memory trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example: See CALCulate<Chn>:DATA
Usage: Query only