SOURce:POWer:CORRection:COLLect:CFACtor <Convergence>
Specifies the convergence factor for a source power calibration.
Parameters:
<Convergence> Convergence factor
Range: 0 to 2
Increment: 0.1
*RST: 0.8
Example: See SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation: See "Flatness Cal – Convergence" on page 526
SOURce:POWer:CORRection:COLLect:FLATness <Boolean>
Enables or disables the source power calibration. The command is valid for all chan-
nels, ports and external generators.
It is not possible to disable flatness calibration and the reference receiver calibration
simultaneously (see SOURce:POWer:CORRection:COLLect:RRECeiver
on page 1350).
Parameters:
<arg0> <Boolean>
Disables (OFF|0) or enables (ON|1) the flatness calibration. With
disabled flatness calibration, only one calibration sweep is per-
formed in order to calibrate the reference receiver; the previous
source calibration data is not overwritten.
*RST: ON|1
Example:
*RST; SOUR:POW:CORR:COLL:VER 0
Disable the verification sweep to speed up the source power cal-
ibration procedure.
SOUR:POW:CORR:COLL:FLAT OFF; RREC ON
Disable flatness cal and enable reference receiver calibration.
SOUR:POW:CORR:ACQ PORT,1
Perform a single source power calibration sweep using port 1
as a source port.
SOURce:POWer:CORRection:COLLect:METHod <PowerCalMethod>
Selects the source power calibration method.
Parameters:
<PowerCalMethod> PMONly | RRAFter | RRONly