Command reference
R&S
®
ZNB/ZNBT
1180User Manual 1173.9163.02 ─ 62
The signals are switched on as soon as a measurement in a segment with non-
zero segment bits is started. They are changed whenever a segment with different
segment bits is measured.
The signals at the USER PORT connector are maintained after the analyzer enters
the hold state. This happens in single sweep mode after all sweep sequences have
been terminated.
Tip:
You can use the active segment number as a parameter for
OUTPut<Ch>:UPORt:SEGMent<Seg>[:VALue] and monitor the measurement in up
to 16 different segments per channel at the USER PORT connector; see example
below. You can also use the USER PORT output signals as segment-dependent trigger
signals for external devices. Use CONTrol:AUXiliary:C[:DATA] to transfer the
four bit value in decimal representation.
Suffix:
<Ch>
.
Channel number.
<Seg> Sweep segment number
Parameters:
<BinValue> Binary value. The transferred values correspond to the following
states of the USER CONTROL connector:
#B0000 - no signal at any of the four pins 16, 17, 18, 19
#B0001 - output signal at pin 16
#B0010 - output signal at pin 17
#B0011 - output signal at pin 16 and 17
...
#B1111 - output signal at pin 16, 17, 18 and 19
Range: #B0000 to #B1111 (for setting command), 0 to 15
(query)
*RST: #B0000 (0)
SCPI command reference