Command reference
R&S
®
ZNB/ZNBT
1222User Manual 1173.9163.02 ─ 62
FRTRans
Transmission normalization, bidirectional
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
FTRans | RTRans
Transmission normalization, forward | reverse
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
REFL
Reflection normalization, Open
RSHort
Reflection normalization, Short
PFNPort
Full n-port (FNPort) calibration with additional power flatness
calibration, for SMARTerCal
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
If an empty string (' ') is specified, the factory cal kit file
stored in the active calibration unit is used.
A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port number. For an n-port automatic calibration, n arbitrary
(not necessarily consecutive) port numbers must be specified.
For a one path two port calibration (OPTPort), the first port no.
denotes the source port (fully calibrated port).
<CalUnitPort1> Port numbers of the cal unit. For an n-port automatic calibration,
n arbitrary (not necessarily consecutive) port numbers must be
specified. It is possible to combine arbitrary (not necessarily
matching) pairs of analyzer and cal unit ports.
<TestPort2> Second test port number
*RST: n/a
Example:
CORR:COLL:AUTO:PORT FNPort, '', 1, 2, 2, 4, 4,
1
Perform an automatic full 3-port calibration at test ports 1, 2, and
4 using the calibration unit's default calibration kit file and ports
2, 4, and 1 of the cal unit.
See also [SENSe<Ch>:]CORRection:COLLect:AUTO:
POWer
Usage: Setting only
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals