Command reference
R&S
®
ZNB/ZNBT
1399User Manual 1173.9163.02 ─ 62
Parameters:
<PmtrName> Name of the external power meter (string parameter). An empty
string means that no particular name is assigned to the power
meter.
<Driver> Power meter type (string parameter). The power meter type is
identical with the name of the power meter driver file (*.pwm)
stored in the resources\extdev subdirectory of the analyz-
er's program directory. Type the driver file name as shown in the
"Add External Powermeter" dialog, i.e. without the file extension
".pwm" (example: use 'NRVD', if the corresponding driver file
name is NRVD.pwm).
Alternative: The '<Driver>' string can also contain the power
meter driver file name with its complete path, e.g. 'C:\Program
Files\Rohde-Schwarz\Vector Network
Analyzer\ZNB\resources\extdev\nrvd.pwm'.
<Interface>
Interface type (string parameter): 'VXI-11', 'socket',
'GPIB0', ..., 'GPIB9', 'other'
<Address> Interface address (string parameter), depending on the interface
type. See Table 5-8
<SensorId> This optional string parameter is used for power meters that can
control several power sensors, such as the R&S NRP2. By
specifying the suitable <SensorId> you can address the related
power sensor.
The R&S NRP2, for example, has four power sensor connec-
tors, Sensor A to Sensor D. By specifying 'A', 'B', 'C' or
'D', you can take control over the power sensor that is attached
to this connector.
Example:
SYST:COMM:RDEV:PMET:CONF:AUTO OFF
Disables "Auto Config NRP-Zxx" (if it was enabled previously).
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter',
'NRP-Z55', 'usb', '100045'
Configure an R&S NRP power meter as external power meter
no. 1, assigning the name "USB Power Meter" and a serial num-
ber 100045.
SYST:COMM:RDEV:PMET:DEL
Clear the power meter configuration table.
SYST:COMM:RDEV:PMET:DEF?
Query the power meter configuration. The analyzer returns an
error message because the power meter no. 1 is no longer con-
figured.
Manual operation:
See "Configured Devices" on page 782
SYSTem:COMMunicate:RDEVice:PMETer:DELete
Clears the configuration table for external power meters.
SCPI command reference