Command reference
R&S
®
ZNB/ZNBT
1420User Manual 1173.9163.02 ─ 62
SYSTem:KLOCk <Boolean>
Locks or unlocks the local controls of the analyzer. This includes the front panel keys,
the keyboard, or other local interfaces.
Parameters:
<Boolean> ON | OFF - lock or unlock the local keys.
*RST: OFF
Example:
SYST:KLOCk ON
Lock the local keys.
SYSTem:LANGuage
<Language>
Specifies the remote language for the analyzer.
Parameters:
<Language> Command syntax for the analyzer, string variable:
'SCPI' R&S ZNB/ZNBT-specific command set: the analyzer
supports all commands described in this documentation.
'ZVR' | 'ZVABT' compatibility with network analyzers of the
R&S ZVR and R&S ZVA/B/T families.
'PNA' | 'HP8510' | 'HP8720' | 'HP8753' | 'HP8714' | 'HP8530' |
'ENA' compatibility with network analyzers from other manu-
facturers.
*RST: n/a - a reset does not affect the language setting.
The factory setting is SCPI.
Example:
SYST:LANG 'PNA'
Select a PNA-compatible command set.
Manual operation: See "Remote Language" on page 777
SYSTem:LOGGing:REMote[:STATe] <Boolean>
Enables logging of all remote control commands transferred to the analyzer.
Parameters:
<Boolean> ON command sequence stored to file
C:\Users\Public\Documents\Rohde-Schwarz\Vna\
RemoteLog\VNARemote.log.
OFF command sequence not logged.
Example:
SYST:LOGG:REM ON
Enable remote logging. The log file
'C:\Users\Public\Documents\Rohde-Schwarz\Vna\
RemoteLog\VNARemote.log' is fixed and does not have to
be specified.