EasyManuals Logo

R&S ZNB Series User Manual

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #590 background imageLoading...
Page #590 background image
GUI reference
R&S
®
ZNB/ZNBT
590User Manual 1173.9163.02 ─ 62
Remote command:
SOURce<Ch>:POWer:CORRection:IMODulation:PORT
Power
Power of the lower and of the upper tone. Both powers are always kept equal. For
power sweeps, the start and stop power must be set in the Frequencies and Power
dialog.
Remote command:
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
Tone Distance
Defines the frequency difference between the upper and the lower tone. The tone dis-
tance is kept constant across the entire sweep.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:TDIStance
Prepare Measurement of IM Order
Selects the orders of the measured intermodulation products. Each order requires an
additional measurement at a different frequency, therefore the measurement time
increases with the number of selected orders. If a higher-order IM is selected, its fre-
quency can exceed the upper or lower frequency limit of the analyzer. Use the "Fit Fre-
quency Range" button in the Frequencies and Power dialog to restrict the sweep range
accordingly.
Remote command:
[SENSe<Ch>:]FREQuency:IMODulation:ORDer<Im>[:STATe]
Step Attenuator
Defines the value for the receiver step attenuator at the receiving port. Same setting as
in the "Power" tab of the "Stimulus" softtool; see "Step Attenuators" on page 417.
Remote command:
[SENSe<Ch>:]POWer:ATTenuation
Meas Bandwidth
Defines the bandwidth of the IF filter at the receiving port.
The setting is identical with the "Bandwidth" setting on the "Bandwidth" tab of the
"Power Bw Avg" softtool; see Chapter 5.9.2, "Bandwidth tab", on page 420.
Remote command:
[SENSe<Ch>:]BWIDth[:RESolution]
Selectivity
Defines the selectivity of the IF filter of the receiving port.
The setting is identical with the "Selectivity" setting on the "Bandwidth" tab of the
"Power Bw Avg" softtool; see Chapter 5.9.2, "Bandwidth tab", on page 420.
Remote command:
[SENSe<Ch>:]BWIDth[:RESolution]:SELect
Channel Config softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals