Command reference
R&S
®
ZNB/ZNBT
995User Manual 1173.9163.02 ─ 62
See also CALCulate<Chn>:TRANsform:DTFault:STARt and CALCulate<Chn>:
TRANsform:DTFault:STOP.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Center>/<Span> Center/span value.
Default unit: m
Example:
:CALCulate1:TRANsform:DTFault:CENTer 2; SPAN 2
Defines the DtF distance window of the active trace of channel 1
via its center (2m) and span (2m).
Options: R&S ZNB/ZNBT-K3
Manual operation: See "Start Distance / Stop Distance" on page 729
CALCulate<Chn>:TRANsform:DTFault:DEFine
<CblName>, <CblPermittivity>,
<CblAtt1>, <CblFreq1>{, <CblAtt2>, <CblFreq2>}...
Defines a new user-defined cable type for Distance to Fault (DtF) measurements.
The cable's frequency-dependent attenuation is defined via attenuation/frequency pairs
<CblAtt1>, <CblFreq1>, <CblAtt2>, <CblFreq2>, ... . At least one pair
has to be specified.
Suffix:
<Chn>
.
Channel number
This suffix is ignored: cable types are defined for all channels.
Setting parameters:
<CblName> Name of the cable type.
Must be unique among all DtF cable types, with case insensitive
match.
<CblPermittivity> Relative permittivity ε
r
of the cable type.
<CblAtt[i]> Attenuation value i in the frequency-dependent attenuation table
of the cable type.
Default unit: dB/m
<CblFreq[i]> Frequency i in the frequency-dependent attenuation table of the
cable type.
Default unit: Hz
Example:
CALCulate:TRANsform:DTFault:DEFine 'My cable
type', 1.1, 0.01, 1GHz, 0.015, 2GHz, 0.012,
3GHz
Defines a new cable type "My cable type" with relative permittiv-
ity ε
r
=1.1 and the frequency-dependent attenuation given in the
table below.
Usage: Setting only