EasyManuals Logo

R&S ZNB Series User Manual

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1021 background imageLoading...
Page #1021 background image
Command reference
R&S
®
ZNB/ZNBT
1021User Manual 1173.9163.02 ─ 62
Parameters for setting and query:
<CiruitModel> STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS |
STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Possible circuit models (character data); see Table 7-7.
Example:
*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:EMB:
PAR:C2? CSSL
Create a balanced port and query the default capacitance C2 for
the Serial Cs, shunt L circuit model. The response is 1E-012 (1
pF).
CALC:TRAN:VNET:BAL:EMB:PAR:C2 CSSL, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation: See "Network" on page 645
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:DATA<Port> <Interchange>, <arg1>
Defines an embedding network for a balanced port based on the given S-Parameter
traces.
Circuit models STSL | STSC | SLST | SCST require S-Parameter traces of two 2-
port networks, to be assigned to the different ports; the FIMPort model requires S-
Parameter traces of a single 4-port network but no additional port assignment.
Use
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
TNDefinition to select the adequate circuit model before executing this com-
mand.
MMEMory:LOAD:VNETworks<Ch>:BALanced:EMBedding<LogPt> to load cir-
cuit data from a Touchstone file located at the R&S ZNB/ZNBT's file system.
Suffix:
<Ch>
.
Channel number
<LogPt> Logical port number
<Port> Port assignment for two 2-port networks:
1 - Port 1
2 - Port 2
This suffix is ignored for 4-port networks.
Setting parameters:
<Interchange> FPORts | IPORts | SGATes | SINCreasing
FPORts (or omitted)
Standard port sequence (odd port numbers towards VNA, even
port numbers towards DUT)
IPORts
– Two-port networks: inverted port sequence (network port 2
towards VNA, network port 1 towards DUT)
– Four-port networks: increasing port sequence (low port num-
bers towards VNA, high port numbers towards DUT)
SCPI command reference