Command reference
R&S
®
ZNB/ZNBT
1039User Manual 1173.9163.02 ─ 62
Parameters:
<Conductance> Conductance G<i> for the specified circuit model.
Range: -1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST: 0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel> STSG | SGST | GSSL | LSSG | GSSG | SLGS | SGLS | SGGS
Circuit model whose conductance G<i> shall be set, see
Table 7-7
Manual operation: See "Network" on page 642
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:L<1|2|3> <CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
PARameters:L<1|2|3>? <CircuitModel>
Specifies the inductance value L<i> in the different lumped circuit models for port pair
deembedding.
In the query form, the <Inductance> parameter must be omitted. The command returns
the inductance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<ListId> Index of the affected port pair (see CALCulate<Ch>:
TRANsform:VNETworks:PPAir:DEEMbedding<ListId>:
DEFine)
<1|2|3> Index i of the inductance L<i> in the related lumped circuit
model. If unspecified the numeric suffix is set to 1.
Parameters:
<Inductance> Inductance L<i>
Range: -1 H to 1 H.
Increment: 1 pH (1E-12 H)
*RST: 1 nH (1E-9 H)
Default unit: H
Parameters for setting and query:
<CircuitModel> STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS |
GSSL | LSSG | SLGS | SGLS
Circuit model whose inductance L<i> shall be set, see Table 7-7
Example: See CALCulate<Ch>:TRANsform:VNETworks:PPAir:
DEEMbedding<ListId>[:STATe]
Manual operation: See "Network" on page 642
SCPI command reference