Command reference
R&S
®
ZNB/ZNBT
1225User Manual 1173.9163.02 ─ 62
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType> FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | POPTport | PFNPort
Calibration Type
FNPort
Full n-port (UOSM) calibration
TOSM
Full n-port calibration with characterized Through.
If the selected cal unit characterization does not contain a
Through characterization, the command silently falls back to a
FNPort (UOSM) calibration.
FOPort
Full one-port ("Refl OSM") calibration
OPTPort
"One Path Two Ports" calibration
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
FRTRans
Transmission normalization, bidirectional
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
FTRans | RTRans
Transmission normalization, forward | reverse
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
REFL
Reflection normalization, Open
RSHort
Reflection normalization, Short
POPTport
"One Path Two Ports" (OPTP) calibration with additional power
flatness calibration (SMARTerCal)
PFNPort
Full n-port (FNPort) calibration with additional power flatness
calibration (SMARTerCal)
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
The empty string (' ') refers to the factory calibration of (and
stored on) the active calibration unit.
A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.