Command reference
R&S
®
ZNB/ZNBT
1233User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
:SENSe1:CORRection:COLLect:METHod:DEFine 'Test', TOSM, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected THRough, 1, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected SHORt, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 1
:SENSe1:CORRection:COLLect:ACQuire:SELected OPEN, 2
:SENSe1:CORRection:COLLect:ACQuire:SELected MATCh, 2
:SENSe1:CORRection:COLLect:DISCard
Usage: Event
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire] <StandardType>,
<TestPort1>[, <TestPort2>]...
Starts a fixture compensation sweep in order to acquire measurement data for a test
fixture that has its inner conductor terminated with the selected standards.
Suffix:
<Ch>
.
Channel number of the corrected channel
Setting parameters:
<StandardType> OPEN | SHORt
Terminating standard type: open or short.
<TestPort1>
<TestPort2> Test port numbers. For a fixture compensation, n arbitrary (not
necessarily consecutive) port numbers must be specified.
*RST: -
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals