EasyManua.ls Logo

R&S ZNB Series - Page 1276

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1276User Manual 1173.9163.02 ─ 62
<Denominator> Range: 1, 2, 3 ...
*RST: 1
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Multiplication tab" on page 600
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:RFPort <Port>
Selects a test port as source port for the RF signal.
Suffix:
<Ch>
.
Channel number
Parameters:
<Port> Test port number
Range: 1 to number of test ports
*RST: 1
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "Port selection" on page 599
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:STAGes <Stages>
Selects the number of mixer stages for scalar mixer measurements.
Suffix:
<Ch>
.
Channel number
Parameters:
<Stages> Number of mixer stages
Range: 1 | 2
*RST: 1
Example: See [SENSe<Ch>:]FREQuency:CONVersion:MIXer:
MFFixed
Manual operation: See "2nd Mixer" on page 601
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:TFRequency<Stg>
<IFConversionMode>
Selects the frequency conversion mode of the IF signal.
Suffix:
<Ch>
.
Channel number.
<Stg> Mixer stage (1 or 2; see [SENSe<Ch>:]FREQuency:
CONVersion:MIXer:STAGes.
SCPI command reference