Command reference
R&S
®
ZNB/ZNBT
1313User Manual 1173.9163.02 ─ 62
Parameters:
<Boolean> ON use independent GC settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol.
OFF reset the GC mode in all sweep segments to the GC
mode for unsegmented sweeps, defined via [SENSe<Ch>:
]POWer:GAINcontrol or [SENSe<Ch>:]POWer:
GAINcontrol:GLOBal. ON will not restore the previous val-
ues.
*RST: ON
Example: See [SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol
Manual operation: See "Segmented AGC" on page 580
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
<IntSourcePow>
Defines the power of the internal signal source in sweep segment no. <Seg>. At the
same time, the command activates separate power control in all sweep segments
([SENSe<Ch>:]SEGMent:POWer[:LEVel]:CONTrol).
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<IntSourcePow> Internal source power
Range: -300 dBm to +200 dBm. The usable power range is
frequency-dependent; refer to the data sheet.
*RST: -10 dBm
Default unit: dBm
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus -10 dBm internal source power.
SEGM:POW -20
Decrease the power to -20 dBm.
Manual operation: See "Optional Columns" on page 435
[SENSe<Ch>:]SEGMent:POWer[:LEVel]:CONTrol <Boolean>
Selects common or independent internal source "Power" settings for the sweep seg-
ments.
Suffix:
<Ch>
.
Channel number
SCPI command reference