Command reference
R&S
®
ZNB/ZNBT
1353User Manual 1173.9163.02 ─ 62
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1356
Usage: Query only
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 530
SOURce:POWer:CORRection:TCOefficient:DEFine<ListNo>
<Frequency>,
<TransCoeff>[, <DeEmbedding Position>]
Adds a new frequency and transmission coefficient to the end of the transmission coef-
ficient list. The query returns the frequency and transmission coefficient no. <ListNo>.
Suffix:
<ListNo>
.
List entry no. (only used for queries)
Parameters:
<Frequency> Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range: Stimulus values outside the frequency range of the
analyzer are allowed.
*RST: n/a
Default unit: Hz
<TransCoeff> Transmission coefficient
Range: -300 dB to +200 dB
*RST: n/a
Default unit: dB
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
SCPI command reference