EasyManua.ls Logo

R&S ZNB Series - Page 1354

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1354User Manual 1173.9163.02 ─ 62
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1356
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 530
SOURce:POWer:CORRection:TCOefficient:DELete:ALL [<DeEmbedding Position>]
Clears the transmission coefficient list.
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example:
See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1356
Usage: Setting only
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 530
SOURce:POWer:CORRection:TCOefficient:DELete<ListNo>[:DUMMy]
[<DeEmbedding Position>]
Deletes entry <ListNo> from the transmission coefficient list.
Suffix:
<Ch>
.
Calibrated channel number. This suffix is ignored; the transmis-
sion coefficient settings are valid for all channels.
<PhyPt> Calibrated port number. This suffix is ignored; the transmission
coefficient settings are valid for all sources.
<ListNo> List entry no.
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
SCPI command reference

Table of Contents