Command reference
R&S
®
ZNB/ZNBT
1355User Manual 1173.9163.02 ─ 62
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1356
Manual operation: See "Frequency / Transm. Coefficients: Insert, Delete, Delete
All" on page 530
SOURce:POWer:CORRection:TCOefficient:FEED <Trace>[, <DeEmbedding
Position>]
Selects a trace which provides the points for the transmission coefficient list.
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Setting parameters:
<Trace> String parameter for the trace name, e.g. 'Trc1'. The trace must
exist in the active recall set.
*RST: n/a
Example: See SOURce:POWer:CORRection:TCOefficient[:STATe]
on page 1356
Usage: Setting only
Manual operation: See "Get Trace..." on page 531