Command reference
R&S
®
ZNB/ZNBT
1433User Manual 1173.9163.02 ─ 62
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce............................................................... 1435
TRIGger<Ch>[:SEQuence]:SLOPe................................................................................ 1435
TRIGger<Ch>[:SEQuence]:SOURce..............................................................................1436
TRIGger<Ch>[:SEQuence]:HOLDoff <DelayTime>
Sets a delay time between the trigger event and the start of the measurement ("Trigger
Delay").
Suffix:
<Ch>
.
Channel number.
Parameters:
<DelayTime> Delay time.
Range: 0 s to 13680 s
Increment: 10 ms
*RST: 0 s
Default unit: s
Example:
TRIG:SOUR MAN
Activate external trigger source.
TRIG:HOLD UP
Set a delay time of 10 ms.
Manual operation: See "Delay" on page 439
TRIGger<Ch>[:SEQuence]:LINK <MeasSequence>
Selects the triggered measurement sequence. The identifier for the sequence is a
string variable.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence> Triggered measurement sequence, string variable.
'SWEep' – trigger event starts an entire sweep.
'SEGMent' – trigger event starts a sweep segment, if segmented
frequency sweep is active (see example below). If another
sweep type is active, the trigger event starts an entire sweep.
'POINt' – trigger event starts measurement at the next sweep
point.
'PPOint' – trigger event starts the next partial measurement at
the current or at the next sweep point.
*RST: 'SWEep'
SCPI command reference