EasyManuals Logo

R&S ZNB Series User Manual

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1449 background imageLoading...
Page #1449 background image
Command reference
R&S
®
ZNB/ZNBT
1449User Manual 1173.9163.02 ─ 62
Example: //Return service information
DIAG:HUMS:SERV?
Returns for example:
#3196{"last":"2021-01-19T23:00:00Z","required":
2,"calibration":{"last":
"2021-01-19T23:00:00Z","nextDue":
"2022-01-19T23:00:00Z","nextDueExpired":
false,"recommendedCalibrationInterval":
"P1Y"},"uptime":0}
Usage: Query only
DIAGnostic:HUMS:STATe
<State>
Turns the HUMS service and data collection on and off.
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Turn on HUMS service
DIAG:HUMS:STAT ON
Manual operation: See "State (Device Utilization)/Status" on page 764
DIAGnostic:HUMS:STORage?
Queries the storage information of the instrument. Depending on the set data format,
the queried data is either displayed in XML or JSON format. For more information about
setting the data format, see DIAGnostic:HUMS:FORMat on page 1447.
Return values:
<StorageInfo> <block_data>
Returns storage information for each storage device of the con-
nected instrument as block data.
#blockdata[{storagedevice1},
{storagedevice2},...]
Binary block data with the following syntax:
#<Digits><Length><Binarydata>
#
Indicates the start of the binary block
<Digits>
Decimal value
Gives the number of decimal digits used for the <Length>
value
<Length>
Decimal value
Number of bytes the follow in the <Binary data> part
<Binary data>
Binary data in ASCII format
For each storage device it returns:
HUMS and service date commands

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals