EasyManua.ls Logo

R&S ZNB Series - Page 1485

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1485User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]SWEep:SPACing <StimulusFreq>
Defines the frequency vs. time characteristics of a frequency sweep ("Lin Frequency"
or "Log Frequency"). The command has no effect on segmented frequency, power or
time sweeps.
Note: Use [SENSe<Ch>:]SWEep:TYPE to select sweep types other than "Lin Fre-
quency" or "Log Frequency".
Suffix:
<Ch>
.
Channel number
Parameters:
<StimulusFreq> LINear | LOGarithmic
The stimulus frequency is swept in equidistant steps over the
frequency range. In a Cartesian diagram, the x-axis is a linear
frequency axis.
Example:
FUNC "XFR:POW:S12"
Activate a frequency sweep and select the S-parameter S
12
as
measured parameter for channel and trace no. 1.
SWE:SPAC LOG
Change to sweep type "Log Frequency".
[SENSe<Chn>:]FUNCtion[:ON] <SweepType>[, <arg1>[, <arg2>]]
Defines the sweep type and the measurement parameter in a single string.
Note: To select a measurement parameter without changing the sweep type, use
CALCulate<Ch>:PARameter:MEASure. Use the other commands in the
CALCulate<Ch>:PARameter... subsystem to create or delete traces and select
measurement parameters.
Suffix:
<Chn>
.
Channel number used to identify the active trace. If
[SENSe<Chn>:]FUNCtion[:ON] is not used as a query, the
number must be 1.
Parameters:
<SweepType> Single string parameter defining the sweep type and the param-
eter to be measured:
<string> = "<sweep_type>:<parameter>".
Range: See list of strings below.
*RST: "XFR:POW:S21"
<arg1> B1 | B2 | A1 | A2 | ABSa1 | ABSa2 | DCIN1 | DCIN2
<arg2> B1 | B2 | A1 | A2 | ABSa1 | ABSa2 | DCIN1 | DCIN2
R&S ZVR/ZVABT compatible commands

Table of Contents