GUI reference
R&S
®
ZNB/ZNBT
388User Manual 1173.9163.02 ─ 62
The appearance of the limit fail symbols is defined in the "Define User Color Scheme"
dialog (see Chapter 5.17.3.2, "Define User Color Scheme dialog", on page 743). You
can choose between various options:
●
Change the trace color between failed measurement points.
●
Show or hide the colored squares.
Note:
●
Circle limit check and display of limit circles are independent of each other:
– The limit circles can be displayed, no matter if the circle limit check is enabled.
– If "Limit Check" is enabled, the ripple limits are checked, no matter if they are
displayed.
●
If result evaluation is limited to a user-defined display circle, the Chapter 5.6.3,
"Circle Test tab", on page 386 is only performed inside this display circle.
●
The circle limits can only be checked if the trace format is complex. While a carte-
sian format is active, the limit circles are hidden and the circle limit check is sus-
pended.
●
If the limit check for (cartesian) limit lines is enabled, it is also evaluated in complex
diagrams.
Figure 5-7: Simultaneous dB Mag limit line and circle check
Remote command:
CALCulate<Chn>:LIMit:CIRCle[:STATe]
CALCulate<Chn>:LIMit:CIRCle:FAIL?
CALCulate:LIMit:CIRCle:FAIL:ALL?
Clear Test
Resets the limit check results.
Remote command:
CALCulate<Chn>:LIMit:CIRCle:CLEar
Draw Circle
Activates touchscreen or mouse operation; tap the diagram at one border of the limit
circle and draw the circle to the required size and position.
Remote command:
n/a
Lines softtool