GUI reference
R&S
®
ZNB/ZNBT
469User Manual 1173.9163.02 ─ 62
Figure 5-24: Calibration Setting Wizard, Step 3: Calibration
In the upper part of the "Calibration" screen, the R&S ZNB/ZNBT shows the sweep dia-
grams for the currently measured S-parameter. The lower part displays the calibrated
ports and standards and visualizes the measurement progress.
Calibration Sweep Diagrams
During the calibration sweep, each diagram contains a single S-parameter trace and a
typical result trace for the measured calibration standard.
If switch matrices are involved, a sweep is performed for every possible signal path
and for each of these paths a separate S-parameter trace is shown (see Chap-
ter 4.7.29.5, "Multiple paths and calibration", on page 275).