GUI reference
R&S
®
ZNB/ZNBT
499User Manual 1173.9163.02 ─ 62
Ports
Selects the test ports to be calibrated.
It is possible to select any combination of two or more test ports. If you are only inter-
ested in a single port p, perform a two-port SMARTerCal for a port pair including p (as
Power Port).
Note: Calibration and port de-/activation.
The analyzer fimware automatically activates/deactivates ports during/after a (success-
ful) calibration:
Calibrated ports that were previously disabled, are automatically enabled as single-
ended logical ports.
An uncalibrated port that is not used by a measurement (i.e. the port is not required
by any trace of the related channel) is disabled.
Remote command:
The port parameters in many calibration commands define ports to be calibrated.
Type
Selects the calibration type to be performed. For background information, see Chap-
ter 4.5.7.2, "Calibration types", on page 200.
SMARTerCal is based on a "One Path Two Ports" or a full n-port calibration (n > 1):
"Adapter Removal" is a 2-port calibration, all other calibration types can be used for
any combination of 2 or more ports.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See "Configured Devices" on page 782.
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Power Port
Selects the port for the scalar source power calibration.
Source leveling
If checked, the "SMARTerCal" wizard opens the Power Cal wizard on completion.
There you can perform a complementary source flatness calibration at the ""SMARTer-
Cal"ibrated" ports, and additional scalar power calibrations at other ports.
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
Proceeds to "Step 2: Connectors and Cal Kits" on page 500. This button is inactive
(and a warning is displayed) if fewer than two ports are selected.
Cal softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals