CALCulate<Chn>:MARKer<Mk>:SEARch:FORMat <SearchFormat>
CALCulate<Chn>:MARKer<Mk>:REFerence:SEARch:FORMat <SearchFormat>
Defines the format in which you want to specify the target value CALCulate<Chn>:
MARKer<Mk>:TARGet for the related marker (see "Marker addressing" on page 940).
Each marker may have a different target format. The table below gives an overview on
how a complex target value z = x + jy is converted.
Target Format Description Formula
MLINear Magnitude of z, unconverted.
|z| = sqrt ( x
2
+ y
2
)
MLOGarithmic Magnitude of z in dB Mag(z) = 20 log|z| dB
PHASe Phase of z φ (z) = arctan (y/x)
UPHase Unwrapped phase of z comprising
the complete number of 360°
phase rotations
Ф(z) = φ (z) + 2k·360°
REAL Real part of z Re(z) = x
IMAGinary Imaginary part of z Im(z) = y
SWR (Voltage) Standing Wave Ratio SWR = (1 + |z|) / (1 – |z|)
DEFault Identical to trace format.
Note: the Smith and Polar traces
use "Lin Mag" as the default for-
mat for target value.
-
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number (ignored for reference markers)
Parameters:
<SearchFormat> MLINear | MLOGarithmic | PHASe | UPHase | REAL |
IMAGinary | SWR | DEFault
The format of the target value in the marker target search
(CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute TAR
Get or CALCulate<Chn>:MARKer<Mk>:REFerence:
FUNCtion:EXECute TARGet). See table above.
*RST: DEFault