EasyManua.ls Logo

R&S ZNB Series - Page 1084

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1084User Manual 1173.9163.02 ─ 62
Parameters:
<Logic> POSitive | NEGative
POSitive 0 = low, 1 = high
NEGative 0 = high, 1 = low
Note:*RST or "Preset" do not change the configuration of the
Universal Interface. Use CONTrol:HANDler:RESet to restore
default values.
*RST: n/a (default: NEGative)
Example:
CONTrol:HANDler:LOGic POS
Change the logic of the data ports to positive.
CONTrol:HANDler:OUTPut<Pt>[:DATA] <BinValue>
Writes a 0 or 1 to the output ports (pin 3 or 4) of the Universal Interface connector. The
port lines have negative logic: A "0" corresponds to a high signal, a "1" to a low signal.
The query reads the last value that has been written to the output port.
Suffix:
<Pt>
.
Output port number:
1 - /OUTPUT1 (pin 3)
2 - /OUTPUT2 (pin 4)
The parameters MIN, MAX, DEF are not supported.
Note:*RST or "Preset" do not change the configuration of the
Universal Interface. Use CONTrol:HANDler:RESet to restore
default values.
Parameters:
<BinValue> 0 - high
1 - low
*RST: n/a (default: 0)
Example:
CONT:HAND:OUTP2:DATA 0
Set the /OUTPUT2 line (pin 4) to 0 (current state of /
OUTPUT2).
CONT:HAND:OUTP2:USER 1
Define the next state of the /OUTPUT2 line as 1 (low). /
OUTPUT2 will go from 0 to 1 when the analyzer receives a neg-
ative pulse on the /INPUT1 line (pin 2).
CONTrol:HANDler:OUTPut<Pt>:USER <BinValue>
Defines the state of the output ports (pin 3 or 4) of the Universal Interface connector
after the next negative pulse on the /INPUT1 line (pin 2).
SCPI command reference

Table of Contents

Related product manuals