Command reference
R&S
®
ZNB/ZNBT
1147User Manual 1173.9163.02 ─ 62
Example:
MMEM:LOAD:CKIT:SDAT 'N 50 Ohm','Default
Kit',MOPEN,'Test data','test.s1p',1
Load the file Test.s1p from the current directory in order to
define the properties of an Open (m) standard in the cal kit
named "Default Kit" for the N 50 Ω connector type. Assign the
label "Test data" and specify that the standard data is valid for
port no. 1 only.
Usage: Setting only
Manual operation: See "Read .s1p File... / Read .s2p File..." on page 518
Connector and cal kit naming conventions:
Connector and calibration kit names must be entered as string parameters. The strings
contain the connector and cal kit names used in the Calibration Kits dialog; a Ω
in the name must be replaced by 'Ohm', e.g.:
'NewKit1' denotes the user-defined calibration kit "NewKit1".
'N 50 Ohm Ideal Kit' denotes the "N 50 Ω Ideal Kit".
'ZV-Z21 typical' denotes the cal kit "ZV-Z21 typical".
MMEMory:LOAD:CKIT:SDATa:WLABel <ConnectorType>, <CalKitName>,
<KitLabel>, <StandardType>, <StandardLabel>, <TouchstoneFile>[,
<FirstPort>[, <SecondPort>]]
Loads characterization data from the given Touchstone file; similar to existing com-
mand MMEMory:LOAD:CKIT:SDATa but supports cal kit addressing by label.
Setting parameters:
<ConnectorType> String parameter containing the name of the connector type.
<CalKitName> String parameter containing the name of a calibration kit availa-
ble on the analyzer.
<KitLabel> String parameter containing the label of a calibration kit available
on the analyzer. An empty string means that no label is defined.
<StandardType> MMTHrough | MFTHrough | FFTHrough | MMLine | MMLine1 |
MMLine2 | MMLine3 | MFLine | MFLine1 | MFLine2 | MFLine3 |
FFLine | FFLine1 | FFLine2 | FFLine3 | MMATten | MFATten |
FFATten | MMSNetwork | MFSNetwork | FFSNetwork | MOPen |
FOPen | MSHort | FSHort | MOSHort | MOSHort1 | MOSHort2 |
MOSHort3 | FOSHort | FOSHort1 | FOSHort2 | FOSHort3 |
MREFlect | FREFlect | MMTCh | FMTCh | MSMatch | FSMatch
Standard types; for a description refer to table Standard types
and their parameters.
<StandardLabel> String parameter addressing a particular calibration standard by
its label. An empty string means that no label is defined.
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals