Command reference
R&S
®
ZNB/ZNBT
1172User Manual 1173.9163.02 ─ 62
<SweepSegFile> String parameter to specify the name and directory of the cre-
ated sweep segment file. The default extension (manual control)
for sweep segment files is *.SegList, although other exten-
sions are allowed.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory
Example: See MMEMory:LOAD:SEGMent
Usage: Setting only
Manual operation: See "Import.../ Export..." on page 434
MMEMory:STORe:STATe
<Compatibility>, <RecallSetFile>
Stores the configuration data of the current recall set to the specified recall set file.
MMEMory:STORe:STATe also renames the current recall set. See example for
MMEMory:LOAD:STATe.
Setting parameters:
<Compatibility> 1 (this value is used for compatibility with the SCPI standard but
ignored).
<RecallSetFile> String parameter to specify the absolute or relative path of the
created recall set file. Relative paths are evaluated relative to
the current directory (see MMEMory:CDIRectory).
The default extension for recall set files is znx. Unless you
specify znxml as file name extension, recall sets are always
stored in znx file format.
Example: See MMEMory:LOAD:STATe
Usage: Setting only
Manual operation: See "Save" on page 671
MMEMory:STORe:TRACe <TraceName>, <TraceFile>[, <FormatInd>[, <Format>[,
<DecSeparator>[, <FieldSeparator>]]]]
Stores the trace data of a specified data trace to a trace file. Traces are created using
the CALCulate<Ch>:PARameter:SDEFine command.
Tip:*.s<n>p Touchstone files (<n> = 1, 2, 3, ...) are intended for a complete set of
<n>-port S-parameters. Data export fails if the active channel does not contain the full
set of <n>
2
traces or if the involved ports are not numbered consecutively, starting with
port 1. If the necessary traces are available, '<trc_name>' can be the name of any
of the traces. To create Touchstone files while less than <n>
2
single-ended traces are
available, use MMEMory:STORe:TRACe:PORTs.
Setting parameters:
<TraceName> Name of an existing data trace in the active recall set (string
parameter).
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals