Command reference
R&S
®
ZNB/ZNBT
1211User Manual 1173.9163.02 ─ 62
Usage: Query only
Manual operation: See "Port Assignment (manual)" on page 454
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine
<TestPort1>, <CalUnitPort1>[, <TestPort2>]...
Manually defines port assignment no. <Asg> for channel no. <Ch>.
With manual configuration a non-minimal set of port assignments can be created,
which may increase measurement accuracy. On the other hand, it is up to the user to
ensure that the created set of port assignments is complete and valid for the chosen
calibration type (see [SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure).
Note that during the corresponding calibration sweep the R&S ZNB/ZNBT expects the
physical port connections to be established exactly as defined by the port pairs.
Use
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:DEFine:
TPORt command to take advantage of auto-detection of port connections.
●
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
or [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:
TPORt:DEFault to create the default assignments instead.
See Chapter 4.5.5.6, "Multiple port assignments", on page 191 for background infor-
mation.
Suffix:
<Ch>
.
Number of the channel
<Asg> Number of the port assignment
Parameters:
<TestPort1> First test port number
<CalUnitPort1> Number of the calibration unit port that is assigned to the first
test port
<TestPort2> Second test port number ...
SCPI command reference