Command reference
R&S
®
ZNB/ZNBT
1210User Manual 1173.9163.02 ─ 62
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See Chapter 4.5.5.6, "Multiple port assignments", on page 191 for background
information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg> Number of the port assignment
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine
Usage: Event
Manual operation: See "Start Cal Sweep / Abort Sweep" on page 457
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:ALL:COUNt?
Returns the total number of port assignments of all calibrations.
See Chapter 4.5.5.6, "Multiple port assignments", on page 191 for background infor-
mation.
Note:
Use [SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt? to get the
number of port assignments of the "current" calibration.
Suffix:
<Ch>
.
Channel number
Example: See Chapter 8.2.5.2, "MultiCal (with calibration unit)",
on page 1516
Usage: Query only
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
Returns the number of port assignments of the "current" calibration, i.e. the calibration
last created using [SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure.
This is particulary useful if the default port assignments were established using
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault.
See Chapter 4.5.5.6, "Multiple port assignments", on page 191 for background infor-
mation.
Note:
In MultiCal scenarios, use [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment:ALL:COUNt? to get the number of port assignments of all calibrations.
Suffix:
<Ch>
.
Channel number
Example: See [SENSe<Ch>:]CORRection:COLLect:AUTO:
ASSignment<Asg>:DEFine on page 1211