EasyManua.ls Logo

R&S ZNB Series - Page 1209

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1209User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]CORRection:COLLect:AUTO <Characterization>, <TestPort1>[,
<TestPort2>[...]]
Selects and initiates an automatic calibration for the specified test ports using a single,
auto-detected port assignment.
Tip:
If the test setup contains a high attenuation, the analyzer may fail to detect the cal
unit ports connected to each of its ports. In this case use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs.
Use [SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE if you want to specify a
particular calibration type for the automatic calibration.
If several calibration units are connected, use SYSTem:COMMunicate:RDEVice:
AKAL:ADDRess to select a unit for the calibration.
Use [SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure to prepare an
automatic calibration with multiple port assignments.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<Characterization> Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
If an empty string (' ') is specified, the factory cal kit file
stored in the active calibration unit is used.
A VNA cal kit file name *.calkit without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>,
<TestPort2>, ...
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied. The analyzer automatically detects the calibration unit ports
connected to each analyzer port.
Example:
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port calibration at test ports 1, 2, and 4
using the calibration unit's default calibration kit file and arbitrary
test ports of the cal unit.
Usage: Setting only
Manual operation: See "Apply/Cancel" on page 458
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
Starts the automatic calibration sweep for the indicated channel and port assignment.
SCPI command reference

Table of Contents

Related product manuals