Command reference
R&S
®
ZNB/ZNBT
1219User Manual 1173.9163.02 ─ 62
– A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
– A VNA cal kit file name *.calkit with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
Note that for inline calibration systems this setting is ignored.
The suitable, ICU-specific characterizations must be specified
using SYSTem:COMMunicate:RDEVice:AKAL:CKIT.
Setting parameters:
<CalName>
Example:
SENSe1:CORRection:COLLect:AUTO:CONFigure
FNPort,''
Sets the automatic calibration to "Full n-port" with factory char-
acterization.
Manual operation: See "Characterization" on page 451
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt <PortNumber>
Defines the "Common Port" of an automatic calibration.
This port will be part of every port assignment that is defined using [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault, or [SENSe<Ch>:
]CORRection:COLLect:AUTO:ASSignment:DEFine:TPORt:DEFault
on page 1214.
If [SENSe:]CORRection:COLLect:AUTO:MCONnect is ON, this port will be used as
the center of all Through measurements. If you set up the port assignments manually,
make sure that every port assignment contains it.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PortNumber> Number of the common port
Usage: Setting only
Manual operation: See "Common Port" on page 452
[SENSe:]CORRection:COLLect:AUTO:MCONnect <MinimumConnect>
Enables or disables the Reduced Through logic for every port assignment of an auto-
matic calibration.
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using [SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
on page 1219.
Setting parameters:
<MinimumConnect> ON (1), or OFF (0)
Manual operation: See "Use Reduced Number of Through" on page 754
SCPI command reference