Command reference
R&S
®
ZNB/ZNBT
1255User Manual 1173.9163.02 ─ 62
Example:
*RST; :CORR:EDEL:AUTO ONCE
Reset the instrument and apply the auto length function to the
default trace (Trc1 in channel 1).
Usage: Setting only
Manual operation: See "Auto Length" on page 618
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:DIELectric <Permittivity>
Defines the permittivity for the offset correction at test port <PhyPt>.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<Permittivity> Permittivity
Range: 1 to +1E+6
*RST: 1.00062
Example: See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation: See "Permittivity / Velocity Factor" on page 617
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:DISTance <MechLength>
Defines the offset parameter for test port <PhyPt> as a mechanical length.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
Parameters:
<MechLength> Mechanical length
Range: -3.402823466E+038 m to +3.4028234664E+038 m.
*RST: 0 m
Default unit: m
Example: See [SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation: See "Delay / Electrical Length / Mech. Length" on page 616
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth <ElecLength>
Defines the offset parameter for test port <PhyPt> as an electrical length.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer
SCPI command reference