EasyManua.ls Logo

R&S ZNB Series - Page 1260

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1260User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>[:STATe] <Boolean>
Resets the offset parameters for all test ports to zero and the reference frequency to 1
GHz or queries whether any of the offset parameters are different from zero.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer. This numeric suffix is ignored; the
command affects the parameters of all ports.
Parameters:
<Boolean> The parameter function depends on whether the command is
used as a setting command or as a query: For setting command:
ON - no effect
OFF - resets all length offsets to zero and the reference fre-
quency to 1 GHz
For query:
1 - at least one length offset parameter is different from its
default value
0 - all length offsets are zero / set to default
*RST: OFF
Example:
*RST; :CORR:OFFS?
Reset the instrument and query whether the length offset param-
eters have been reset as well. The response is 0.
Manual operation: See "Reset Offsets" on page 607
[SENSe<Ch>:]CORRection:PCAL <ActivatePowerCals>
Activates or deactivates all power calibrations.
Suffix:
<Ch>
.
Calibrated channel number
Setting parameters:
<ActivatePowerCals> NONE | ALL
NONE deactivate all power calibrations.
ALL activate all power calibrations.
*RST: n/a
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is 1).
CORR:PCAL NONE; POW:AWAV?
Disable all power calibrations and check again whether the cali-
bration is applied (the response is 0).
Usage: Setting only
SCPI command reference

Table of Contents

Related product manuals