Command reference
R&S
®
ZNB/ZNBT
1259User Manual 1173.9163.02 ─ 62
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> OFF (0)
f
2
not used (default)
ON (1)
f
2
used
Manual operation: See "Loss at DC / Loss at 1st Freq / 1st Freq for Loss / Use 2nd
Freq / Loss at 2nd Freq / 2nd Freq for Loss" on page 607
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:COMPensation[:STATe]
<Boolean>
Toggles length/loss/fixture compensation for physical port <PhyPt> ON/OFF.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> ON (1): compensation active
OFF (0): compensation inactive
Manual operation: See "Active" on page 606
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Returns whether a direct fixture compensation has been carried out at port no.
<PhyPt>. A direct fixture compensation resets the offset parameters to zero, the ana-
lyzer uses calculated transmission factors instead.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt> Port number of the analyzer.
Return values:
<Boolean> 1 - direct fixture compensation data used
0 - no direct fixture compensation data used
Example:
*RST; CORR:OFFS:DFC?
Reset the instrument and query whether the analyzer uses direct
fixture compensation data at port 1. The response is 0.
Usage: Query only
Manual operation: See "Direct Compensation" on page 620