Command reference
R&S
®
ZNB/ZNBT
1280User Manual 1173.9163.02 ─ 62
Parameters:
<PortNumber> Analyzer port number
Range: 1 to the number of analyzer ports
*RST: 2
Example: See [SENSe<Ch>:]FREQuency:IMODulation:LTONe
Manual operation: See "Receiving Port" on page 589
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum:MORDer <IMOrder>
Defines the maximum order of intermodulation products for the intermodulation spec-
trum measurement.
Suffix:
<Ch>
.
Channel number
Parameters:
<IMOrder> Maximum order of IM products
Range: 3 | 5 | 7 | 9
*RST: 3
Example: See [SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:
STATe]
Manual operation: See "CW Mode Spectrum" on page 583
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:STATe] <Boolean>
Enables or disables the measurement of the intermodulation spectrum. If a new chan-
nel is desired, the trace must be created separately; see second example below.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON | OFF - enables or disables intermodulation spectrum mea-
surement.
*RST: OFF
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals