Command reference
R&S
®
ZNB/ZNBT
1287User Manual 1173.9163.02 ─ 62
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Permittivity / Velocity Factor" on page 688
[SENSe<Ch>:]HARMonic:RTIMe:DATA <DUTRiseTime>
Sets/gets the minimum rise time the user wishes to measure on the DUT.
Use [SENSe<Ch>:]HARMonic:RTIMe:THReshold to select the underlying rise time
definition.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseTime> Default unit: s
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Rise Time" on page 688
[SENSe<Ch>:]HARMonic:RTIMe:THReshold <DUTRiseThreshold>
Defines how the rise time that is set using [SENSe<Ch>:]HARMonic:RTIMe:DATA
shall be interpreted. Allows to select between rise time definitions 20%-80% (default)
and 10%-90%.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseThreshold> T1_9 | T2_8
Rise time definition:
T1_9: 10%-90%
T2_8: 20%-80%
*RST: T2_8
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Rise Time Definition" on page 688
[SENSe<Ch>:]HARMonic:VELocity:DATA <DUTVelocity>
Sets/gets the velocoty factor of the DUT.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.