Command reference
R&S
®
ZNB/ZNBT
1309User Manual 1173.9163.02 ─ 62
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect <TimeRef>
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the
command [SENSe<Ch>:]SEGMent<Seg>:INSert, is entered as a segment sweep
time ("Segment Time") or as a measurement delay ("Meas Delay").
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<TimeRef> SWTime | DWELl
SWTime - use segment sweep time.
DWELl - use measurement delay.
*RST: SWTime
Example:
SEGM1:INS:SEL DWEL
Select the meas. delay to determine the sweep time in a new
sweep segment no. 1.
SEGM1:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz and a meas. delay of 10 ms.
SEGM1:SWE:TIME?
Query the sweep time in the new segment.
Manual operation: See "Optional Columns" on page 435
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>:POWer[:LEVel] <IntSourcePortPow>
Defines a port-specific power level for port <PortId> in sweep segment <Seg> of chan-
nel <Ch>.
Requires that a port-specific power level has been previously enabled using
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>[:STATe].
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
<PortId> Port number
Parameters:
<IntSourcePortPow> Power level for port <PortId> in sweep segment <Seg> of chan-
nel <Ch>
Default unit: dBm
Manual operation: See "Optional Columns" on page 435