Command reference
R&S
®
ZNB/ZNBT
1308User Manual 1173.9163.02 ─ 62
<SegmentTime> Duration of the sweep in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:SWEep:TIME. In the setting [SENSe<Ch>:
]SEGMent<Seg>:INSert:SELect DWELl, this parameter is
replaced by <MeasDelay>.
Range: Depending on other channel settings. AUTO acti-
vates automatic sweep time setting in the segment,
which is equivalent to the minimum sweep time
possible.
Default unit: s
<MeasDelay> Delay for each partial measurement in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime,
this parameter is replaced by <SegmentTime>.
<Unused> Ignored parameter (for compatibility with R&S ZVR analyzers).
Should be set to the default value 0.
<MeasBandwidth> IF bandwidth in the segment. See [SENSe<Ch>:
]SEGMent<Seg>:BWIDth[:RESolution].
Default unit: Hz
<LO> POSitive | NEGative | AUTO
Position of the local oscillator frequency LO relative to the RF
frequency. In remote control this parameter must be set when a
sweep segment is created.
See [SENSe<Ch>:]FREQuency:SBANd.
<Selectivity> NORMal | MEDium | HIGH
Selectivity of the IF filter. See [SENSe<Ch>:]SEGMent<Seg>:
BWIDth[:RESolution]:SELect.
<Mode> AUTO
Ignored.
<FreqSweepMode> STEPped | ANALog
"Stepped"mode (STEPped, default) or "Swept" mode
(=ANALog).
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Manual operation: See "Add / Insert / Delete / Delete All" on page 433
SCPI command reference