[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl <MeasDelay>
Defines the delay time for each partial measurement in sweep segment no. <Seg>
("Meas. Delay"). If coupling of the segments is switched on ([SENSe<Ch>:
]SEGMent<Seg>:SWEep:DWELl:CONTrol ON), the delay is valid for all sweep seg-
ments in the current channel.
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<MeasDelay> Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep ([SENSe<Ch>:
]SEGMent<Seg>:SWEep:TIME).
Range: 0 s to 2500 s
*RST: 0 s
Default unit: s
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:SWE:DWEL 1 MS
Set the meas. delay in segment no. 1 to 1 ms.
SEGM:DEF? Response:
300000,8000000000,51,-300,0.056559,0,10000,POS,
NORM
Query the channel parameters for sweep segment 1. The
response value for the segment sweep time (olive) implicitly con-
tains the defined meas. delay.
Manual operation: See "Optional Columns" on page 435
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl:CONTrol <Boolean>
Selects common or independent "Meas. Delay" settings for the sweep segments.
Suffix:
<Ch>
.
Channel number.
<Seg> Sweep segment number. This suffix is ignored; the command
controls the whole segmented sweep.
SCPI command reference