Command reference
R&S
®
ZNB/ZNBT
1390User Manual 1173.9163.02 ─ 62
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
Queries all characterizations (cal kit files) which are stored on the connected calibra-
tion unit, either on the calibration unit's internal (flash) memory or on an SD card inser-
ted at the calibration unit (prefix: "SD"). A possible response is
'Factory,ZN-Z51_custom,Throughs,SD:test'. The factory characterization is
always available; an empty string denotes that no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZN-Z51::1234'
Select the cal unit named 'ZN-Z51::1234'.
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
Query all characterizations stored on the connected calibration
unit.
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
'My_calkit', OPEN, S11, 1
Query the characterization data for the characterization named
'My_calkit' and an open standard (one-port, port restriction). A
characterization with the queried properties must be available on
the cal unit.
SYSTem:COMMunicate:RDEVice:AKAL:WARMup?
Query the warmup status of the calibration unit.
SYSTem:COMMunicate:RDEVice:AKAL:DATE? 'Factory'
Query the creation date of the factory calibration.
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
'Factory'
Query the frequency range of the factory calibration.
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
'Factory'
Query the port assignment of the factory calibration.
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:
CATalog? 'Factory'
Query the standards of the factory calibration. Possible
response:
'MOP(P1),MSH(P1),MMMT(P1),MOP(P2),MSH(P2),
MMMT(P2)' denotes an Open (m), Short (m) and Match (m)
standard at each of the ports 1 and 2.
Usage:
Query only
Manual operation: See "Characterization" on page 451
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals