SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog? <CalKitName>
Queries all calibration standards of a characterization (calkit file) that is stored on a
connected calibration unit, either on the calibration unit's internal (flash) memory or on
an SD card inserted at the calibration unit (if available). The query SYSTem:
COMMunicate:RDEVice:AKAL:CKIT:CATalog? returns the names of the cal kit
files. The factory characterization is always available.
Query parameters:
<CalKitName> Name of the calkit file, string parameter.
'Factory' denotes the factory characterization, the prefix
"SD:" indicates that the characterization is stored on an SD card
inserted at the calibration unit.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 451
SYSTem:COMMunicate:RDEVice:AKAL:DATE? <CalKitName>
Queries the creation date and time of the cal unit characterization (calkit file) <CalKit-
Name>. A possible response is 'Friday, May 26, 2011, 10:13:40'. An empty
string is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
Query parameters:
CalKitName String parameter containing the name of a cal unit characteriza-
tion (calkit file).
'Factory' denotes the factory characterization; an empty
string ' ' refers to the last referenced characterization; the pre-
fix "SD:" indicates that the characterization is stored on an SD
card inserted at the calibration unit.
Example: See SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
CATalog?
Usage: Query only
Manual operation: See "Characterization" on page 451
SYSTem:COMMunicate:RDEVice:AKAL:FRANge? <CalKitName>
Queries the frequency range of the cal unit characterization (calkit file) <CalKitName>.
The response consists of the start and stop frequencies in Hz, separated by a comma.
0,0 is returned if no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via SYSTem:COMMunicate:RDEVice:AKAL:ADDRess.
SCPI command reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals