Command reference
R&S
®
ZNB/ZNBT
1396User Manual 1173.9163.02 ─ 62
Example:
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02',
'gpib0', '21'
Configure an R&S SME02 generator as external generator no.
1, assigning the name "Ext. Gen. 1" and a GPIB address 21.
SYST:COMM:RDEV:GEN1:DEF?
Query the generator configuration. The response 'Ext. Gen.
1','SME02','GPIB0','21',OFF, OFF indicates that the
fast sweep mode is disabled and that the analyzer is set to inter-
nal reference frequency.
SYST:COMM:RDEV:GEN1:DEL
Clear the generator configuration table.
SYST:COMM:RDEV:GEN1:DEF?
Query the generator configuration. The analyzer returns an error
message because the generator no. 1 is no longer configured.
Manual operation:
See "Configured Devices" on page 787
SYSTem:COMMunicate:RDEVice:GENerator:DELete
Clears the configuration table for external generators.
Example:
See SYSTem:COMMunicate:RDEVice:GENerator<Gen>:
DEFine
Usage: Event
Manual operation: See "Configured Devices" on page 787
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPMode <Type>
Defines the external generator power control after the sweep end.
Suffix:
<Gen>
.
Generator number
Parameters:
<Type> OFF | LOW | KEEP | USER
OFF power is switched off
LOW switch to min. power
KEEP keep power at the value selected via SOURce<Ch>:
POWer<PhyPt>:GENerator<Gen>:OFFSet
USER set to end power selected via SYSTem:COMMunicate:
RDEVice:GENerator<Gen>:SEPower
Example: See SYSTem:COMMunicate:RDEVice:GENerator<Gen>:
SEPower
Manual operation: See "Configured Devices" on page 787
SCPI command reference