Command reference
R&S
®
ZNB/ZNBT
1415User Manual 1173.9163.02 ─ 62
Tip: Switching off the display speeds up the measurement. This command may have
an impact on the update of trace and channel settings; see SYSTem:SETTings:
UPDate.
Parameters:
<Activate> ON | OFF - switch the display on or off. If the display is switched
on, the analyzer shows the diagrams and traces like in manual
control.
ONCE - switch the display on and show the current trace. This
parameter can be used for occasional checks of the measure-
ment results or settings. The measurement is continued, how-
ever, the measurement results are not updated. Compared to
the ON setting, ONCE does not slow down the measurement
speed.
Example:
SYST:DISP:UPD ON
Switch the display on to view the traces and diagrams.
SYSTem:DISPlay:XLABels
<Boolean>
Enables or disables the display of X-axis grid labels in cartesian diagrams with linear
scale.
Parameters:
<Boolean> ON (1)
Show X-axis grid labels
OFF (0)
Don't show X-axis grid labels
*RST: OFF
Manual operation: See "Show X-Axis Grid Labels" on page 756
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entries consist of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section Error messages and troubleshooting.
Example:
SYST:ERR:ALL?
Query all entries in the error queue. 0,"No error" is returned
if the error queue is empty.
Usage: Query only
SYSTem:ERRor:COUNt?
Gets the current number of entries in the SCPI error queue.
Usage:
Query only
SCPI command reference