EasyManua.ls Logo

R&S ZNB Series - Page 1437

R&S ZNB Series
1611 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNB/ZNBT
1437User Manual 1173.9163.02 ─ 62
sists of TTL pulses of configurable duration and polarity which can be sent before or
after each sweep or measurement point.
Figure 7-1: Output trigger signal at EXT TRIG OUT
TRIGger:CHANnel<Ch>:AUXiliary:DURation.................................................................. 1437
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]................................................................... 1437
TRIGger:CHANnel<Ch>:AUXiliary:INTerval.....................................................................1438
TRIGger:CHANnel<Ch>:AUXiliary:OPOLarity................................................................. 1438
TRIGger:CHANnel<Ch>:AUXiliary:POSition....................................................................1438
TRIGger:CHANnel<Ch>:AUXiliary:DURation <TrigOutDuration>
Specifies the width of the output trigger pulses at Trigger Out.
The trigger duration must be shorter than the selected sweep interval (see TRIGger:
CHANnel<Ch>:AUXiliary:INTerval).
If the trigger duration is too long, you can modify one of the following settings:
Reduce the trigger duration, if possible.
Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
Parameters:
<TrigOutDuration> Pulse width between 1 μs (1E-6 s) and 1 s.
*RST: 1 us
Default unit: s
Example: See TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]
on page 1437
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle] <EnableTriggOut>
Turns the external trigger output EXT TRIG OUT for channel no. <Ch> on or off.
Suffix:
<Ch>
.
Channel number
SCPI command reference

Table of Contents

Related product manuals