Command reference
R&S
®
ZNB/ZNBT
1437User Manual 1173.9163.02 ─ 62
sists of TTL pulses of configurable duration and polarity which can be sent before or
after each sweep or measurement point.
Figure 7-1: Output trigger signal at EXT TRIG OUT
TRIGger:CHANnel<Ch>:AUXiliary:DURation.................................................................. 1437
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]................................................................... 1437
TRIGger:CHANnel<Ch>:AUXiliary:INTerval.....................................................................1438
TRIGger:CHANnel<Ch>:AUXiliary:OPOLarity................................................................. 1438
TRIGger:CHANnel<Ch>:AUXiliary:POSition....................................................................1438
TRIGger:CHANnel<Ch>:AUXiliary:DURation <TrigOutDuration>
Specifies the width of the output trigger pulses at Trigger Out.
The trigger duration must be shorter than the selected sweep interval (see TRIGger:
CHANnel<Ch>:AUXiliary:INTerval).
If the trigger duration is too long, you can modify one of the following settings:
●
Reduce the trigger duration, if possible.
●
Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
Parameters:
<TrigOutDuration> Pulse width between 1 μs (1E-6 s) and 1 s.
*RST: 1 us
Default unit: s
Example: See TRIGger:CHANnel<Ch>:AUXiliary[:ENABle]
on page 1437
TRIGger:CHANnel<Ch>:AUXiliary[:ENABle] <EnableTriggOut>
Turns the external trigger output EXT TRIG OUT for channel no. <Ch> on or off.
Suffix:
<Ch>
.
Channel number
SCPI command reference