Command reference
R&S
®
ZNB/ZNBT
1470User Manual 1173.9163.02 ─ 62
<Mk> Marker number.
Ignored for reference markers and for bandfilter searches
(BFILter), because bandfilter searches always use markers
M1 to M4.
Parameters:
<Mode> MAXimum | MINimum | RPEak | LPEak | NPEak | TARGet |
LTARget | RTARget | BFILter | MMAXimum | MMINimum |
SPRogress
See CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
CALCulate<Chn>:MARKer<Mk>:FUNCtion:TARGet <SearchValue>
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:TARGet <SearchValue>
Defines the target value for a target search with the related marker (see "Marker
addressing" on page 940), which can be activated using CALCulate<Chn>:
MARKer<Mk>:FUNCtion:EXECute TARGet or CALCulate<Chn>:MARKer<Mk>:
REFerence:FUNCtion:EXECute TARGet.
Note: These commands are the ZVR-compatible equivalents of CALCulate<Chn>:
MARKer<Mk>:TARGet and CALCulate<Chn>:MARKer<Mk>:REFerence:TARGet.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk> Marker number (ignored for reference markers)
Parameters:
<SearchValue> Target search value of marker no. <Mk>.
Range: Depending on the format of the active trace. For a
dB Mag trace the range is -200 dB to +200 dB.
Increment: 0.1 dB
*RST: Depending on the trace format; 0 dB for a dB Mag
trace.
Default unit: dB
CALCulate<Chn>:MARKer<Mk>:MAXimum
CALCulate<Chn>:MARKer<Mk>:MINimum
CALCulate<Chn>:MARKer<Mk>:REFerence:MAXimum
CALCulate<Chn>:MARKer<Mk>:REFerence:MINimum
These commands select a search mode for the related marker (see "Marker address-
ing" on page 940) and initiate a maximum or minimum search.
The marker must be created before using CALCulate<Chn>:MARKer<Mk>[:STATe]
ON or CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON.
Note: These commands are the ZVR-compatible equivalents of CALCulate<Chn>:
MARKer<Mk>:FUNCtion:EXECute MINimum | MAXimum and CALCulate<Chn>:
MARKer<Mk>:REFerence:FUNCtion:EXECute MINimum | MAXimum.
R&S ZVR/ZVABT compatible commands