GUI reference
R&S
®
ZNB/ZNBT
443User Manual 1173.9163.02 ─ 62
Restart Sweep
Stops the current measurement cycle and starts a new one.
For the DEFAULT Remote Language, the effect of this command depends on the
sweep modes of the active channel, and the restart behavior defined in the Restart
Manager dialog:
"Sweep Active Channel" "Sweep All Channels"
Active channel in "Continuous"
sweep mode
The "Continuous" sweep mode
channels are swept cyclically, the
"Single" sweep channels are on
hold.
First the "Single" sweep mode
channels are swept one after the
other, each with its channel-spe-
cific number of "Sweeps". Then
the "Continuous" sweep mode
channels are sweep cyclically.
Active channel in "Single"
sweep mode
First the active channel is swept
with its channel-specific number
of "Sweeps". Then the "Continu-
ous" sweep mode channels are
sweep cyclically.
For remote languages other than DEFAULT, the effect of the "Restart Sweep" com-
mand is determined by the selected (common) sweep mode and restart behavior:
"Sweep Active Channel" "Sweep All Channels"
All channels in "Continuous"
sweep mode
The channels are swept cyclically
All channels in "Single" sweep
mode
The active channel is swept
"Sweeps" times. All other chan-
nels are on hold
The channels are swept one after
the other, each of them "Sweeps"
times.
Remote command:
INITiate<Ch>[:IMMediate][:DUMMy]
INITiate[:IMMediate]:ALL
Restart Manager...
Opens the Restart Manager dialog.
All Channels Continuous/All Channels on Hold
Selects a common sweep mode for all channels of the active recall set.
"All Channels Continuous": all channels are set to "Continuous" sweep mode
"All Channels on Hold": all channels are set to "Single" sweep mode.
Restart Sweep initiates a new measurement cycle.
These actions are only available for the DEFAULT Remote Language). For other
remote languages, the Continuous / Single setting applies to all channels and the "All
Channels Continuous"/"All Channels on Hold" buttons are greyed out;
Remote command:
INITiate:CONTinuous:ALL
Sweep Softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals