Command reference
R&S
®
ZNB/ZNBT
1133User Manual 1173.9163.02 ─ 62
Example: Alternative settings using the R&S ZVAB compatibility mode:
*RST; :SYSTem:LANGuage 'ZVABT'
*RST; :INIT:CONT OFF
Activate single sweep mode for all channels (including channel
no. 2 created later).
INIT:SCOP SING
State that a single sweep will be performed in the active chan-
nel.
CALC2:PAR:SDEF 'TRC2','S11'; :INIT2
Create channel no. 2 with a new trace and start a single sweep
in channel no. 2. Start a single sweep in the second channel.
Manual operation: See "Continuous / Single" on page 442
INITiate:CONTinuous:ALL <Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Parameters:
<Boolean> ON - the analyzer measures continuously in all channel, repeat-
ing the current sweep. The query returns ON (1) if at least one
channel is measured continuously.
OFF - the measurement is stopped after the number of sweeps
defined via [SENSe<Ch>:]SWEep:COUNt. INITiate<Ch>[:
IMMediate][:DUMMy] initiates a new measurement cycle.
*RST: ON
Example: See INITiate<Ch>:CONTinuous
Manual operation: See "All Channels Continuous/All Channels on Hold"
on page 443
INITiate[:IMMediate]:ALL
Starts a new single sweep sequence in all channels. This command is available in sin-
gle sweep mode only (INITiate:CONTinuous:ALL OFF). The data of the last sweep
(or of previous sweeps, see Retrieving the results of previous sweeps) can be read
using CALCulate<Chn>:DATA:NSWeep:FIRSt? SDATa, <count>.
This command is not supported in the "ZVR" or "ZVABT" compatibility modes.
Note: In contrast to all other commands of the analyzer, the
INITiate<Ch>[:IMMediate]... commands have been implemented for overlap-
ped execution; see Chapter 6.4, "Command processing", on page 821.
Example:
See INITiate<Ch>:CONTinuous
Usage: Event
Manual operation: See "Restart Sweep" on page 443
SCPI command reference