648User Manual 1173.9163.02 ─ 62
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:R
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:G
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:C
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:L
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:R
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
PARameters:G
Select File...
"Select File..." is enabled as long as the "1-Port Data" network is selected. This net-
work is defined by its S-parameters stored in a one-port Touchstone file (*.s1p). No
additional parameters are required.
Note: The loaded file is stored in the active recall set. Persisted recall sets contain the
full (de-)embedding data so that they can be transferred to other instruments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:GLOop:DEEMbedding<group>
MMEMory:LOAD:VNETworks<Ch>:GLOop:EMBedding<group>
Set to Ideal Ground
This function is enabled as long as the 1-Port Data network is active. An imported S-
parameter set is replaced by the S-parameters of an ideal through connection, which
eliminates the transformation network.
Ground Loop per Port Group
If selected, each port group (= DUT) can have its own ground loop deembedding and
embedding networks. Otherwise the same de-/embedding network is used for all active
ports.
The checkbox is only visible if the instrument supports multiple DUTs, which is the case
for R&S ZNBT and 4-port R&S ZNB with 2 internal sources.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:GROup
Offset Embed softtool

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNB Series and is the answer not in the manual?

R&S ZNB Series Specifications

General IconGeneral
BrandR&S
ModelZNB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals