Parameters:
<Conductance> Conductance G for the specified circuit model.
Range: -1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST: 0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel> SG
Possible circuit models (character data); see Table 7-9.
Manual operation: See "Network" on page 647
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L <CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:L? <CircuitModel>
Specifies the inductance value in the different circuit models for ground loop deembed-
ding.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<Inductance> Inductance L for ground loop deembedding.
Range: -1 H to 1 H.
Increment: 1 pH (1E-12 H)
*RST: 1 nH (1E-9 H)
Default unit: H
Parameters for setting and query:
<CircuitModel> SL
Possible circuit models (character data); see Table 7-9.
Example:
*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:L? SL
Query the default inductance for ground loop deembedding. The
response is 1E-009 (1 nH).
CALC:TRAN:VNET:GLO:DEEM:PAR:L SL, 2.2E-9
Increase the inductance to 2.2 nH.
Manual operation: See "Network" on page 647